JAJSEG5A January   2018  – October 2018 ADS112U04

PRODUCTION DATA.  

  1. 特長
  2. アプリケーション
  3. 概要
    1.     Device Images
      1.      Kタイプ熱電対温度の測定
  4. 改訂履歴
  5. 概要(続き)
  6. Pin Configuration and Functions
    1.     Pin Functions
  7. Specifications
    1. 7.1 Absolute Maximum Ratings
    2. 7.2 ESD Ratings
    3. 7.3 Recommended Operating Conditions
    4. 7.4 Thermal Information
    5. 7.5 Electrical Characteristics
    6. 7.6 UART Timing Requirements
    7. 7.7 UART Switching Characteristics
    8. 7.8 Typical Characteristics
  8. Parameter Measurement Information
    1. 8.1 Noise Performance
  9. Detailed Description
    1. 9.1 Overview
    2. 9.2 Functional Block Diagram
    3. 9.3 Feature Description
      1. 9.3.1  Multiplexer
      2. 9.3.2  Low-Noise Programmable Gain Stage
        1. 9.3.2.1 PGA Input Voltage Requirements
        2. 9.3.2.2 Bypassing the PGA
      3. 9.3.3  Voltage Reference
      4. 9.3.4  Modulator and Internal Oscillator
      5. 9.3.5  Digital Filter
      6. 9.3.6  Conversion Times
      7. 9.3.7  Excitation Current Sources
      8. 9.3.8  Sensor Detection
      9. 9.3.9  System Monitor
      10. 9.3.10 Temperature Sensor
        1. 9.3.10.1 Converting From Temperature to Digital Codes
          1. 9.3.10.1.1 For Positive Temperatures (For Example, 50°C):
          2. 9.3.10.1.2 For Negative Temperatures (For Example, –25°C):
        2. 9.3.10.2 Converting From Digital Codes to Temperature
      11. 9.3.11 Offset Calibration
      12. 9.3.12 Conversion Data Counter
      13. 9.3.13 Data Integrity
      14. 9.3.14 General-Purpose Digital Inputs/Outputs
    4. 9.4 Device Functional Modes
      1. 9.4.1 Power-Up and Reset
        1. 9.4.1.1 Power-On Reset
        2. 9.4.1.2 RESET Pin
        3. 9.4.1.3 Reset by Command
      2. 9.4.2 Conversion Modes
        1. 9.4.2.1 Single-Shot Conversion Mode
        2. 9.4.2.2 Continuous Conversion Mode
      3. 9.4.3 Operating Modes
        1. 9.4.3.1 Normal Mode
        2. 9.4.3.2 Turbo Mode
        3. 9.4.3.3 Power-Down Mode
    5. 9.5 Programming
      1. 9.5.1 UART Interface
        1. 9.5.1.1 Receive (RX)
        2. 9.5.1.2 Transmit (TX)
        3. 9.5.1.3 Data Ready (DRDY)
        4. 9.5.1.4 Protocol
        5. 9.5.1.5 Timeout
      2. 9.5.2 Data Format
      3. 9.5.3 Commands
        1. 9.5.3.1 RESET (0000 011x)
        2. 9.5.3.2 START/SYNC (0000 100x)
        3. 9.5.3.3 POWERDOWN (0000 001x)
        4. 9.5.3.4 RDATA (0001 xxxx)
        5. 9.5.3.5 RREG (0010 rrrx)
        6. 9.5.3.6 WREG (0100 rrrx dddd dddd)
        7. 9.5.3.7 Command Latching
      4. 9.5.4 Reading Data
        1. 9.5.4.1 Manual Data Read Mode
        2. 9.5.4.2 Automatic Data Read Mode
      5. 9.5.5 Data Integrity
    6. 9.6 Register Map
      1. 9.6.1 Configuration Registers
      2. 9.6.2 Register Descriptions
        1. 9.6.2.1 Configuration Register 0 (address = 00h) [reset = 00h]
          1. Table 18. Configuration Register 0 Field Descriptions
        2. 9.6.2.2 Configuration Register 1 (address = 01h) [reset = 00h]
          1. Table 19. Configuration Register 1 Field Descriptions
        3. 9.6.2.3 Configuration Register 2 (address = 02h) [reset = 00h]
          1. Table 21. Configuration Register 2 Field Descriptions
        4. 9.6.2.4 Configuration Register 3 (address = 03h) [reset = 00h]
          1. Table 22. Configuration Register 3 Field Descriptions
        5. 9.6.2.5 Configuration Register 4 (address = 04h) [reset = 00h]
          1. Table 23. Configuration Register 4 Field Descriptions
  10. 10Application and Implementation
    1. 10.1 Application Information
      1. 10.1.1 Interface Connections
      2. 10.1.2 Analog Input Filtering
      3. 10.1.3 External Reference and Ratiometric Measurements
      4. 10.1.4 Establishing Proper Limits on the Absolute Input Voltage
      5. 10.1.5 Unused Inputs and Outputs
      6. 10.1.6 Pseudo Code Example
    2. 10.2 Typical Applications
      1. 10.2.1 K-Type Thermocouple Measurement (–200°C to +1250°C)
        1. 10.2.1.1 Design Requirements
        2. 10.2.1.2 Detailed Design Procedure
        3. 10.2.1.3 Application Curves
      2. 10.2.2 3-Wire RTD Measurement (–200°C to +850°C)
        1. 10.2.2.1 Design Requirements
        2. 10.2.2.2 Detailed Design Procedure
          1. 10.2.2.2.1 Design Variations for 2-Wire and 4-Wire RTD Measurements
        3. 10.2.2.3 Application Curves
      3. 10.2.3 Resistive Bridge Measurement
        1. 10.2.3.1 Design Requirements
        2. 10.2.3.2 Detailed Design Procedure
  11. 11Power Supply Recommendations
    1. 11.1 Power-Supply Sequencing
    2. 11.2 Power-Supply Decoupling
  12. 12Layout
    1. 12.1 Layout Guidelines
    2. 12.2 Layout Example
  13. 13デバイスおよびドキュメントのサポート
    1. 13.1 ドキュメントのサポート
      1. 13.1.1 関連資料
    2. 13.2 ドキュメントの更新通知を受け取る方法
    3. 13.3 コミュニティ・リソース
    4. 13.4 商標
    5. 13.5 静電気放電に関する注意事項
    6. 13.6 Glossary
  14. 14メカニカル、パッケージ、および注文情報

パッケージ・オプション

メカニカル・データ(パッケージ|ピン)
サーマルパッド・メカニカル・データ
発注情報

Configuration Register 1 (address = 01h) [reset = 00h]

Figure 70. Configuration Register 1
7 6 5 4 3 2 1 0
DR[2:0] MODE CM VREF[1:0] TS
R/W-0h R/W-0h R/W-0h R/W-0h R/W-0h

Table 19. Configuration Register 1 Field Descriptions

Bit Field Type Reset Description
7:5 DR[2:0] R/W 0h Data rate.
These bits control the data rate setting depending on the selected operating mode. Table 20 lists the bit settings for normal and turbo mode.
4 MODE R/W 0h Operating mode.
These bits control the operating mode that the device operates in.

0 : Normal mode (256-kHz modulator clock, default)
1 : Turbo mode (512-kHz modulator clock)
3 CM R/W 0h Conversion mode.
This bit sets the conversion mode for the device.

0 : Single-shot conversion mode (default)
1 : Continuous conversion mode
2:1 VREF[1:0] R/W 0h Voltage reference selection.
These bits select the voltage reference source that is used for the conversion.

00 : Internal 2.048-V reference selected (default)
01 : External reference selected using the REFP and REFN inputs
10 : Analog supply (AVDD – AVSS) used as reference
11 : Analog supply (AVDD – AVSS) used as reference
0 TS R/W 0h Temperature sensor mode.
This bit enables the internal temperature sensor and puts the device in temperature sensor mode.
The settings of configuration register 0 have no effect and the device uses the internal reference for measurement when temperature sensor mode is enabled.

0 : Temperature sensor mode disabled (default)
1 : Temperature sensor mode enabled

Table 20. DR Bit Settings

NORMAL MODE TURBO MODE
000 = 20 SPS 000 = 40 SPS
001 = 45 SPS 001 = 90 SPS
010 = 90 SPS 010 = 180 SPS
011 = 175 SPS 011 = 350 SPS
100 = 330 SPS 100 = 660 SPS
101 = 600 SPS 101 = 1200 SPS
110 = 1000 SPS 110 = 2000 SPS
111 = Reserved 111 = Reserved