SBAU487 August 2025
The current sensor inputs to the ADS131M08 are located on three-terminal screw connectors J4, J8, J11, and J14. Figure 2-2 depicts the input structure common to all four current sensor inputs.
The shunt jumpers on JP1 through JP8 must be installed to provide a burden resistor when using current transformers. Rogowski coils can also be connected to these inputs in which case the shunt jumpers need to be removed as no burden resistor is needed when using this type of current sensor. Figure 2-3 provides additional detail.