TIDM-EKG-PULSE
低功耗脈搏血氧儀和 EKG 參考設計
TIDM-EKG-PULSE
概覽
This reference design details non-invasive optical plethysmography - pulsoximetry - using the MSP430FG437 microcontroller (MCU). The pulsoximeter consists of a peripheral probe combined with the MCU displaying the oxygen saturation and pulse rate on a LCD glass. The same sensor is used for both heart-rate detection and pulsoximetering in this application. A custom LabVIEW GUI has been developed for this application to display the heart signal, heart rate and % SaO2 number. All the necessary files and steps to get this GUI up and running have been provided along with this report.
特點
- Single-chip solution with integrated analog front-end for diode drive and measurement
- USB communication interface for PC connectivity
- Built-in connectivity for addition of wireless communication using CCxxxx EMK transceivers
- NellcorTM probe compatible hardware interface
- Based on original design "A Single-Chip Pulse oximeter Design Using the MSP430" (SLAA274)
- This application report uses the MOD-PULSE Pulse oximeter and Heart-Rate Monitor using the MSP430FG439 development board by OLIMEX Ltd.
- This reference design is tested and includes software, hardware design files, documentation & PC GUI (LabView)
已開發完全組裝的電路板,僅供測試與性能驗證,且為非賣品。
設計檔案與產品
設計檔案
下載立即可用的系統檔案以加速您的設計流程。
技術文件
找不到結果。請清除您的搜尋條件,然後再試一次。
檢視所有 1
| 類型 | 標題 | 下載最新的英文版本 | 日期 | |||
|---|---|---|---|---|---|---|
| 應用說明 | Revised Pulse Oximeter Design Using the MSP430 | 2010/6/3 |