AFE4420

활성

다중 센서 웨어러블 광학 심박수 모니터링을 위한 FIFO가 포함된 초소형 통합 AFE

제품 상세 정보

Applications Optical-HRM Number of input channels 4 Resolution (bps) 22 Number of LEDs 4 Interface type I2C Operating temperature range (°C) -40 to 85 Rating Catalog
Applications Optical-HRM Number of input channels 4 Resolution (bps) 22 Number of LEDs 4 Interface type I2C Operating temperature range (°C) -40 to 85 Rating Catalog
DSBGA (YZ) 30 4.8 mm² 2.4 x 2
  • Supports signal acquisition of up to 16 phases
  • Flexible allocation of LEDs and PDs in each phase
  • Accurate, continuous heart-rate monitoring:
    • Up to 100-dB dynamic range
    • Low current for continuous operation on a wearable device with a typical value
      • 15 µA for an LED, 20 µA for the receiver
  • Transmitter:
    • Supports common anode LED configuration
    • 8-Bit programmable LED current with range adjustable from 50 mA to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of 4 LEDs for optimized SpO2, HRM, or multi-wavelength HRM
    • Programmable control signals to extend operation to 8 LEDs using external SPDT switches
  • Receiver:
    • Supports 4 time-multiplexed photodiode inputs
    • Individual DC offset subtraction DAC at TIA input for each LED, ambient with 7-bit control and range up to ±254 µA
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Digital ambient subtraction at ADC output to suppress ambient tones
    • Noise filtering with programmable bandwidth
  • Supports external clock and internal oscillator modes
  • Automatic ambient cancellation mode (AACM) to increase receiver dynamic range
  • Two programmable GPIO pins with flexible interrupt generation
  • FIFO with 128-sample depth
  • I2C, SPI interfaces: selectable by pin
  • Operating temperature range: –40°C to +85°C
  • 2.6-mm × 2.1-mm DSBGA, 0.4-mm pitch
  • Supplies:
    • Rx:
      • 1.7 V to 1.9 V (LDO bypass)
      • 1.9 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.5 V
    • IO: 1.7 V to RX_SUP
  • Supports signal acquisition of up to 16 phases
  • Flexible allocation of LEDs and PDs in each phase
  • Accurate, continuous heart-rate monitoring:
    • Up to 100-dB dynamic range
    • Low current for continuous operation on a wearable device with a typical value
      • 15 µA for an LED, 20 µA for the receiver
  • Transmitter:
    • Supports common anode LED configuration
    • 8-Bit programmable LED current with range adjustable from 50 mA to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of 4 LEDs for optimized SpO2, HRM, or multi-wavelength HRM
    • Programmable control signals to extend operation to 8 LEDs using external SPDT switches
  • Receiver:
    • Supports 4 time-multiplexed photodiode inputs
    • Individual DC offset subtraction DAC at TIA input for each LED, ambient with 7-bit control and range up to ±254 µA
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Digital ambient subtraction at ADC output to suppress ambient tones
    • Noise filtering with programmable bandwidth
  • Supports external clock and internal oscillator modes
  • Automatic ambient cancellation mode (AACM) to increase receiver dynamic range
  • Two programmable GPIO pins with flexible interrupt generation
  • FIFO with 128-sample depth
  • I2C, SPI interfaces: selectable by pin
  • Operating temperature range: –40°C to +85°C
  • 2.6-mm × 2.1-mm DSBGA, 0.4-mm pitch
  • Supplies:
    • Rx:
      • 1.7 V to 1.9 V (LDO bypass)
      • 1.9 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.5 V
    • IO: 1.7 V to RX_SUP

The AFE4420 is an analog front-end for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The device supports up to four switching light-emitting diodes (LEDs) and up to four photodiodes. Up to 16 signal phases can be defined and the signal can be acquired from each phase in a synchronized manner and stored in a 128-sample First in, First out (FIFO) block. The FIFO can be read out using either an I2C or a SPI interface. The AFE also has a fully integrated LED driver with an 8-bit current control. The device has a high dynamic range transmit-and-receive circuitry that helps with the sensing of very small signal levels.

The AFE4420 is an analog front-end for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The device supports up to four switching light-emitting diodes (LEDs) and up to four photodiodes. Up to 16 signal phases can be defined and the signal can be acquired from each phase in a synchronized manner and stored in a 128-sample First in, First out (FIFO) block. The FIFO can be read out using either an I2C or a SPI interface. The AFE also has a fully integrated LED driver with an 8-bit current control. The device has a high dynamic range transmit-and-receive circuitry that helps with the sensing of very small signal levels.

다운로드 스크립트와 함께 비디오 보기 동영상
추가 정보

NDA에 따라 전체 데이터시트 및 기타 설계 리소스를 사용할 수 있습니다. 지금 요청

관심 가지실만한 유사 제품

open-in-new 대안 비교
다른 핀 출력을 지원하지만 비교 대상 장치와 동일한 기능
AFE44I30 활성 FIFO 및 I2C를 지원하는 웨어러블 광학 바이오센싱용 초저전력 통합 AFE(아날로그 프론트 엔드) Higher SNR, 6 LEDs, 4 PDs, and improved automatic ambient and LED cancellation.
AFE44S30 활성 FIFO 및 SPI를 지원하는 웨어러블 광학 바이오센싱용 초저전력 통합 AFE(아날로그 프론트 엔드) Higher SNR, 8 LEDs, 4 PDs, and improved automatic ambient and LED cancellation.

기술 자료

star =TI에서 선정한 이 제품의 인기 문서
검색된 결과가 없습니다. 검색어를 지우고 다시 시도하십시오.
3개 모두 보기
유형 직함 날짜
* Data sheet AFE4420 Ultra-Small, Integrated AFE With FIFO for Multi-sensor Wearable, Optical Heart-rate Monitoring and Bio-sensing datasheet (Rev. A) PDF | HTML 2019/07/30
Technical article Designing compact, efficient and high performing multiparameter patient monitors PDF | HTML 2019/06/25
Technical article Going beyond traditional SpO2 measurement with multiwavelength optical measurement PDF | HTML 2019/02/12

설계 및 개발

추가 조건 또는 필수 리소스는 사용 가능한 경우 아래 제목을 클릭하여 세부 정보 페이지를 확인하세요.

평가 보드

AFE4420EVM — 웨어러블 광학 심박수 모니터링을 위한 FIFO가 포함된 AFE4420 통합 AFE 평가 모듈

AFE4420 EVM(평가 모듈)은 HRM(심박수 모니터링) 및 SpO2(주변 모세혈관 산소포화도)와 같은 광학 바이오센싱 애플리케이션을 위한 AFE(아날로그 프론트 엔드)인 AFE4420 장치를 평가하기 위한 플랫폼입니다. AFE4420은 스위칭 LED(발광 다이오드) 4개와 포토다이오드 4개를 지원합니다. 포토다이오드의 전류는 TIA(트랜스임피던스 증폭기)를 통해 전압으로 변환된 후 ADC(아날로그-디지털 컨버터)를 사용하여 디지털화됩니다. ADC 코드는 프로그래밍 가능한 깊이를 지원하는 128샘플 (...)

시뮬레이션 툴

PSPICE-FOR-TI — TI 설계 및 시뮬레이션 툴용 PSpice®

TI용 PSpice®는 아날로그 회로의 기능을 평가하는 데 사용되는 설계 및 시뮬레이션 환경입니다. 완전한 기능을 갖춘 이 설계 및 시뮬레이션 제품군은 Cadence®의 아날로그 분석 엔진을 사용합니다. 무료로 제공되는 TI용 PSpice에는 아날로그 및 전력 포트폴리오뿐 아니라 아날로그 행동 모델에 이르기까지 업계에서 가장 방대한 모델 라이브러리 중 하나가 포함되어 있습니다.

TI 설계 및 시뮬레이션 환경용 PSpice는 기본 제공 라이브러리를 이용해 복잡한 혼합 신호 설계를 시뮬레이션할 수 있습니다. 레이아웃 및 제작에 (...)
레퍼런스 디자인

TIDA-010029 — Bluetooth® 5를 지원하는 웨어러블, 16상 멀티 센서 SpO2 및 심박수 모니터(HRM) 레퍼런스 설계

This reference design enables a wearable, optimized saturation of peripheral capillary oxygen (SpO2) and multi-sensor, multi-wavelength optical heart rate monitor (HRM). It uses AFE4420 device, which is a single-chip, bio-sensing front end for photoplethysmography (PPG) measurements. It supports up (...)
Design guide: PDF
회로도: PDF
패키지 CAD 기호, 풋프린트 및 3D 모델
DSBGA (YZ) 30 Ultra Librarian

주문 및 품질

포함된 정보:
  • RoHS
  • REACH
  • 디바이스 마킹
  • 납 마감/볼 재질
  • MSL 등급/피크 리플로우
  • MTBF/FIT 예측
  • 물질 성분
  • 인증 요약
  • 지속적인 신뢰성 모니터링
포함된 정보:
  • 팹 위치
  • 조립 위치

권장 제품에는 본 TI 제품과 관련된 매개 변수, 평가 모듈 또는 레퍼런스 디자인이 있을 수 있습니다.

지원 및 교육

TI 엔지니어의 기술 지원을 받을 수 있는 TI E2E™ 포럼

콘텐츠는 TI 및 커뮤니티 기고자에 의해 "있는 그대로" 제공되며 TI의 사양으로 간주되지 않습니다. 사용 약관을 참조하십시오.

품질, 패키징, TI에서 주문하는 데 대한 질문이 있다면 TI 지원을 방문하세요. ​​​​​​​​​​​​​​

동영상