AFE4420
- Supports signal acquisition of up to 16 phases
- Flexible allocation of LEDs and PDs in each phase
- Accurate, continuous heart-rate monitoring:
- Up to 100-dB dynamic range
- Low current for continuous operation on a wearable device with a typical value
- 15 µA for an LED, 20 µA for the receiver
- Transmitter:
- Supports common anode LED configuration
- 8-Bit programmable LED current with range adjustable from 50 mA to 200 mA
- Mode to fire two LEDs in parallel
- Programmable LED on-time
- Simultaneous support of 4 LEDs for optimized SpO2, HRM, or multi-wavelength HRM
- Programmable control signals to extend operation to 8 LEDs using external SPDT switches
- Receiver:
- Supports 4 time-multiplexed photodiode inputs
- Individual DC offset subtraction DAC at TIA input for each LED, ambient with 7-bit control and range up to ±254 µA
- Transimpedance gain: 10 kΩ to 2 MΩ
- Digital ambient subtraction at ADC output to suppress ambient tones
- Noise filtering with programmable bandwidth
- Supports external clock and internal oscillator modes
- Automatic ambient cancellation mode (AACM) to increase receiver dynamic range
- Two programmable GPIO pins with flexible interrupt generation
- FIFO with 128-sample depth
- I2C, SPI interfaces: selectable by pin
- Operating temperature range: –40°C to +85°C
- 2.6-mm × 2.1-mm DSBGA, 0.4-mm pitch
- Supplies:
- Rx:
- 1.7 V to 1.9 V (LDO bypass)
- 1.9 V to 3.6 V (LDO enabled)
- Tx: 3 V to 5.5 V
- IO: 1.7 V to RX_SUP
- Rx:
The AFE4420 is an analog front-end for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The device supports up to four switching light-emitting diodes (LEDs) and up to four photodiodes. Up to 16 signal phases can be defined and the signal can be acquired from each phase in a synchronized manner and stored in a 128-sample First in, First out (FIFO) block. The FIFO can be read out using either an I2C or a SPI interface. The AFE also has a fully integrated LED driver with an 8-bit current control. The device has a high dynamic range transmit-and-receive circuitry that helps with the sensing of very small signal levels.
추가 정보
NDA에 따라 전체 데이터시트 및 기타 설계 리소스를 사용할 수 있습니다. 지금 요청
관심 가지실만한 유사 제품
다른 핀 출력을 지원하지만 비교 대상 장치와 동일한 기능
기술 자료
유형 | 직함 | 날짜 | ||
---|---|---|---|---|
* | Data sheet | AFE4420 Ultra-Small, Integrated AFE With FIFO for Multi-sensor Wearable, Optical Heart-rate Monitoring and Bio-sensing datasheet (Rev. A) | PDF | HTML | 2019/07/30 |
Technical article | Designing compact, efficient and high performing multiparameter patient monitors | PDF | HTML | 2019/06/25 | |
Technical article | Going beyond traditional SpO2 measurement with multiwavelength optical measurement | PDF | HTML | 2019/02/12 |
설계 및 개발
추가 조건 또는 필수 리소스는 사용 가능한 경우 아래 제목을 클릭하여 세부 정보 페이지를 확인하세요.
AFE4420EVM — 웨어러블 광학 심박수 모니터링을 위한 FIFO가 포함된 AFE4420 통합 AFE 평가 모듈
AFE4420 EVM(평가 모듈)은 HRM(심박수 모니터링) 및 SpO2(주변 모세혈관 산소포화도)와 같은 광학 바이오센싱 애플리케이션을 위한 AFE(아날로그 프론트 엔드)인 AFE4420 장치를 평가하기 위한 플랫폼입니다. AFE4420은 스위칭 LED(발광 다이오드) 4개와 포토다이오드 4개를 지원합니다. 포토다이오드의 전류는 TIA(트랜스임피던스 증폭기)를 통해 전압으로 변환된 후 ADC(아날로그-디지털 컨버터)를 사용하여 디지털화됩니다. ADC 코드는 프로그래밍 가능한 깊이를 지원하는 128샘플 (...)
PSPICE-FOR-TI — TI 설계 및 시뮬레이션 툴용 PSpice®
TI 설계 및 시뮬레이션 환경용 PSpice는 기본 제공 라이브러리를 이용해 복잡한 혼합 신호 설계를 시뮬레이션할 수 있습니다. 레이아웃 및 제작에 (...)
TIDA-010029 — Bluetooth® 5를 지원하는 웨어러블, 16상 멀티 센서 SpO2 및 심박수 모니터(HRM) 레퍼런스 설계
패키지 | 핀 | CAD 기호, 풋프린트 및 3D 모델 |
---|---|---|
DSBGA (YZ) | 30 | Ultra Librarian |
주문 및 품질
- RoHS
- REACH
- 디바이스 마킹
- 납 마감/볼 재질
- MSL 등급/피크 리플로우
- MTBF/FIT 예측
- 물질 성분
- 인증 요약
- 지속적인 신뢰성 모니터링
- 팹 위치
- 조립 위치
권장 제품에는 본 TI 제품과 관련된 매개 변수, 평가 모듈 또는 레퍼런스 디자인이 있을 수 있습니다.