SNAS602C FEBRUARY 2013 – December 2014 ADS1293
PRODUCTION DATA.
NOTE
Information in the following applications sections is not part of the TI component specification, and TI does not warrant its accuracy or completeness. TI’s customers are responsible for determining suitability of components for their purposes. Customers should validate and test their design implementation to confirm system functionality.
The ADS1293 is an AFE for biopotential measurements. The device is typically used in portable, low-power medical, sports, and fitness ECG applications. The device's flexibility and synchronization features allows it to be used it in configurations that range from single-chip/single-channel applications to multi-chip/multi-channel applications. The following sections will explore the use model in some of these configurations.
In the 3-Lead ECG example shown in Figure 84, the right-arm (RA), left-arm (LA), left-leg (LL) and right-leg (RL) electrodes are connected to the IN1, IN2, IN3 and IN4 pins. The ADS1293 uses the Common-Mode Detector to measure the common-mode of the system by averaging the voltage of input pins IN1, IN2 and IN3, and uses this signal in the right-leg drive feedback circuit
Figure 84. 3-Lead ECG Application
For this design example, use the parameters listed in Table 12 as the set-up parameters.
| DESIGN PARAMETER | EXAMPLE VALUE |
|---|---|
| Number of electrodes | 4 |
| Lead I definition | LA – RA |
| Lead II definition | LL –RA |
| Bandwidth | 175 Hz |
| Output data rate | 853 sps |
| Analog supply voltage | 5.0 V |
| Digital I/O supply voltage | 3.3 V |
Follow the next steps to configure the device for this example, starting from default registers values.
Follow the description in the Streaming section to read the data. The ADS1293 will measure lead I and lead II. Lead III can be calculated as follows: Lead III = Lead II – Lead I
Optionally, the third channel could be used to measure Lead III.
Figure 85 show measurement data collected by a single ADS1293 device connected to an ECG simulator configured to produce an ECG signals at a rate of 60 per minute and with an amplitude of 2 mV. The data was collected simultaneously by channels 1 and 2 of the device during a period of 10 seconds.

Figure 85. Single ADS1293 Device With an ECG Simulator, CH1 and CH2: ECG Signals
In the 5-Lead ECG example shown in Figure 86, the ADS1293 uses the Common-Mode Detector to measure the common-mode of the system by averaging the voltage of input pins IN1, IN2 and IN3, and uses this signal in the right-leg drive feedback circuit
Figure 86. 5-Lead ECG Application
For this design example, use the parameters listed in Table 13 as the setup parameters.
| DESIGN PARAMETER | EXAMPLE VALUE |
|---|---|
| Number of electrodes | 5 |
| Lead I definition | LA – RA |
| Lead II definition | LL –RA |
| Lead V definition | V1 – WCT |
| Bandwidth | 175 Hz |
| Output data rate | 853 sps |
| Analog supply voltage | 5.0 V |
| Digital I/O supply voltage | 3.3 V |
The following steps configure the ADS1293 for a 5-lead application with an ECG bandwidth of 175 Hz and an output data rate of 853 Hz; it is assumed that the device registers contain their default power-up values.
Follow the description in the Streaming section to read the data.
Figure 87 show measurement data collected by a single ADS1293 device connected to an ECG simulator configured to produce an ECG signals at a rate of 60 per minute and with an amplitude of 2 mV. The data was collected simultaneously by channels 1, 2 and 3 of the device during a period of 10 seconds.


Figure 87. Single ADS1293 Device With an ECG Simulator, CH1, CH2, and CH3: ECG Signals
Figure 88. 8- or 12-Lead ECG Application
Figure 88 shows the ADS1293 master/slave setup for an 8-Lead to 12-Lead ECG system. The ADS1293 uses the Common-Mode Detector to measure the common-mode of the system by averaging the voltage of input pins IN1, IN2 and IN3, and uses this signal in the right-leg drive feedback circuit
For this design example, use the parameters listed in Table 14 as the setup parameters.
| DESIGN PARAMETER | EXAMPLE VALUE |
|---|---|
| Number of electrodes | 10 |
| Lead I definition | LA – RA |
| Lead II definition | LL –RA |
| Lead V1 definition | V1 – WCT |
| Lead V2 definition | V2 – WCT |
| Lead V3 definition | V3 – WCT |
| Lead V4 definition | V4 – WCT |
| Lead V5 definition | V5 – WCT |
| Lead V6 definition | V6 – WCT |
| Bandwidth | 175 Hz |
| Output data rate | 853 sps |
| Analog supply voltage | 5.0 V |
| Digital I/O supply voltage | 3.3 V |
The next steps will configure the master device; it is assumed that the device registers contain their default power-up values.
Next, configure the slave devices; it is assumed that the device registers contain their default power-up values. In this example, both devices will have the same configuration; therefore, they can potentially be configured in parallel by asserting the CSB signal of both chips.
Finally, start the conversion. This should be written to all three chips.
The three devices will run synchronously using the SYNCB signal. Follow the description in the Streaming section to read the data. The ADS1293 measures lead I, lead II and leads V1-V6. For a 12-lead application, the remaining 4 leads can be calculated as follows:
Figure 89 show measurement data collected by 3 synchronized ADS1293 devices connected to an ECG simulator configured to produce an ECG signals at a rate of 60 per minute with an amplitude of 2 mV. The data was collected simultaneously by multiple channels from all 3 devices during a period of 10 seconds.







Figure 89. Three Synchornized ADS1293 Devices With an ECG Simulator: ECG Signals