SBOU187A april   2017  – april 2023 INA233

 

  1.   Abstract
  2.   Trademarks
  3. 1Overview
    1. 1.1 INA233EVM Kit Contents
    2. 1.2 Related Documentation from Texas Instruments
  4. 2INA233EVM Hardware
    1. 2.1 Theory of Operation for INA233 Hardware
    2. 2.2 Signal Definitions of H1 (10-Pin Connector Socket)
      1. 2.2.1 SM-USB-DIG Platform Description
  5. 3INA233EVM Hardware Setup
    1. 3.1 Electrostatic Discharge Warning
    2. 3.2 One-Time INA233EVM GUI Software Setup
      1. 3.2.1 Hardware Requirements
      2. 3.2.2 Software Installation
    3. 3.3 Configuration of INA233EVM Jumper Settings
    4. 3.4 Connecting the Hardware
      1. 3.4.1 System Power Up
    5. 3.5 INA233EVM Features
      1. 3.5.1 J3: I2C VS Control Setting
      2. 3.5.2 J1: I2C Address Hardware Setting (A0)
      3. 3.5.3 J2: I2C Address Hardware Setting (A1)
      4. 3.5.4 External I2C Lines and Test Points SCL and SDA
      5. 3.5.5 IN+/IN– Input Filter (R1, R2, and C1)
      6. 3.5.6 Shunt Monitor Configuration
  6. 4INA233EVM Software Overview
    1. 4.1 Starting the INA233EVM Software
    2. 4.2 Configuring the INA233EVM Software
      1. 4.2.1 I2C Address Selection
      2. 4.2.2 Set the Configuration Register (MFR_ADC_CONFIG)
        1. 4.2.2.1 Set Averaging Mode
        2. 4.2.2.2 Set Conversion Times
        3. 4.2.2.3 Set Operating Mode
      3. 4.2.3 Set Calibration Register (MFR_CALIBRATION)
      4. 4.2.4 Set the MFR_DEVICE_CONFIG Register
      5. 4.2.5 Set the MFR_ALERT_MASK
      6. 4.2.6 Set the Warning Limit Registers
    3. 4.3 Using the INA233EVM Software
      1. 4.3.1 Read Data Tab
      2. 4.3.2 Status Register Tab
      3. 4.3.3 Register Map Tab
      4. 4.3.4 Plots Tab
  7. 5INA233EVM Documentation
    1. 5.1 Schematic
    2. 5.2 PCB Layout
    3. 5.3 Bill of Materials
  8. 6Revision History

Set Calibration Register (MFR_CALIBRATION)

The Calibration Register must then be set correctly for the software to operate properly. The user needs to manually calculate the desired value using Equation 1 and then input that value into the MFR_Calibration box, as shown in Figure 4-7. After this, the user must press W in order to write into the Calibration Register. If inputs were correct the MFR_CALIBRATION and MFR_CALIBRATION Hex boxes should display the same values after pressing R (Pressing R reads the Calibration Register).

Equation 1. Calibration Register= 0.00512Current_LSB×RSHUNT

The current LSB is calculated by a recommended range in the INA233 data sheet as shown in Equation 2. It is important to note that the Current LSB and the Calibration Register values are calculated based on the other variable and the RSHUNT value. See the section, Programming the INA233 in the product data sheet for more information on setting the Calibration Register value.

Equation 2. Maximum Expected Current215Current_LSB Maximum Expected Current212
GUID-97E9B84C-5DD8-4551-9670-468FB562821B-low.png Figure 4-7 Setting the Calibration Register (MFR_CALIBRATION)