SLVUDY9 September   2026

 

  1.   1
  2.   Description
  3.   Get Started
  4.   Features
  5.   5
  6. 1Evaluation Module Overview
    1. 1.1 Introduction
    2. 1.2 Kit Contents
    3. 1.3 Specification
    4. 1.4 Device Information
  7. 2Hardware
    1. 2.1 Power Requirements
    2. 2.2 Jumper Information
    3. 2.3 Setup
  8. 3Software
    1. 3.1 GUI Installation
    2. 3.2 Software Description
      1. 3.2.1 DAC90802 Register Page
      2. 3.2.2 DAC90802 DC Output Page
      3. 3.2.3 DC Linearity Page
      4. 3.2.4 Arbitrary Waveform Generation Page
        1. 3.2.4.1 Quick Start
        2. 3.2.4.2 GUI Control Reference
        3. 3.2.4.3 Update Rate
      5. 3.2.5 Multiplying Mode Page
        1. 3.2.5.1 Multiplying Bandwidth Measurement Setup
        2. 3.2.5.2 Multiplying THD Measurement
  9. 4Hardware Design Files
    1. 4.1 Schematics
    2. 4.2 PCB Layouts
    3. 4.3 Bill of Materials (BOM)
  10. 5Compliance Information
    1. 5.1 Compliance and Certifications
  11. 6Related Documentation

Multiplying Bandwidth Measurement Setup

Figure 3-12 illustrates the recommended connections and instructions for multiplying bandwidth measurement using a network analyzer (bode 100 network analyzer). Verify all procedural instructions are completed prior to initiating the measurement:

  1. Calibrate the measuring instruments before connecting to the EVM for measurement.
  2. Connect the network analyzer as shown in Figure 3-12.
  3. Configure the desired frequency sweep in the measuring instrument.
  4. Configure the DAC in multiplying mode by clicking on the Configure tab in the GUI page.
  5. Measurement can now be started.
DAC90802EVM-PDK DAC90802EVMGUI Multiplying Mode Page Figure 3-12 DAC90802EVMGUI Multiplying Mode Page