AFE4900

활성

제품 상세 정보

Applications 1-lead ECG, Optical-HRM Number of input channels 3 Resolution (Bits) 22 Number of LEDs 4 Interface type I2C, SPI Operating temperature range (°C) -20 to 70 Rating Catalog
Applications 1-lead ECG, Optical-HRM Number of input channels 3 Resolution (Bits) 22 Number of LEDs 4 Interface type I2C, SPI Operating temperature range (°C) -20 to 70 Rating Catalog
DSBGA (YZ) 30 4.8 mm² (— mm × — mm)
  • Synchronized PPG, ECG signal acquisition at data rates up to 1 kHz
  • ECG signal chain:
    • Standalone ECG acquisition up to 4 kHz
    • Input bias: 1-lead ECG with RLD bias
    • Programmable INA gain: 2.15 to 12.92
    • Input noise (1 Hz to 150 Hz): 2.5 µVrms at 1 kHz, 1.25 µVrms at 4-kHz data rate
    • AC, DC lead-off detect: 12.5-nA to 100-nA
    • IEC 60601 Test report available on request
  • PPG receiver:
    • Supports three time-multiplexed PD inputs
    • 24-Bit representation of current from PD
    • DC offset subtraction DAC (Up to ±126-µA) at TIA input for each LED, ambient
    • Digital ambient subtraction at ADC output
    • Noise filtering with programmable bandwidth
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Dynamic range up to 100 dB
    • Receiver operates in PPG-only mode at approximately 1-µA/Hz sampling rate
    • Power-down mode: approximately 0 µA
  • PPG transmitter:
    • Four LEDs in common anode configuration
    • 8-Bit LED current up to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of three LEDs for SpO2, or multiwavelength HRM
    • Average current of 30 µA adequate for a
      typical heart-rate monitoring scenario:
      • 20-mA Setting, 60-µs pulse duration,
        25-Hz sampling rate
  • Clocking using an external or internal clock
  • FIFO with 128-sample depth for ECG and PPG
  • I2C, SPI interfaces: selectable by pin
  • 2.6-mm × 2.1-mm, 0.4-mm Pitch DSBGA package
  • Supplies:
    • Rx: 1.8 V to 1.9 V (LDO bypass),
      2.0 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.25 V
    • IO: 1.7 V to Rx_SUP
  • Synchronized PPG, ECG signal acquisition at data rates up to 1 kHz
  • ECG signal chain:
    • Standalone ECG acquisition up to 4 kHz
    • Input bias: 1-lead ECG with RLD bias
    • Programmable INA gain: 2.15 to 12.92
    • Input noise (1 Hz to 150 Hz): 2.5 µVrms at 1 kHz, 1.25 µVrms at 4-kHz data rate
    • AC, DC lead-off detect: 12.5-nA to 100-nA
    • IEC 60601 Test report available on request
  • PPG receiver:
    • Supports three time-multiplexed PD inputs
    • 24-Bit representation of current from PD
    • DC offset subtraction DAC (Up to ±126-µA) at TIA input for each LED, ambient
    • Digital ambient subtraction at ADC output
    • Noise filtering with programmable bandwidth
    • Transimpedance gain: 10 kΩ to 2 MΩ
    • Dynamic range up to 100 dB
    • Receiver operates in PPG-only mode at approximately 1-µA/Hz sampling rate
    • Power-down mode: approximately 0 µA
  • PPG transmitter:
    • Four LEDs in common anode configuration
    • 8-Bit LED current up to 200 mA
    • Mode to fire two LEDs in parallel
    • Programmable LED on-time
    • Simultaneous support of three LEDs for SpO2, or multiwavelength HRM
    • Average current of 30 µA adequate for a
      typical heart-rate monitoring scenario:
      • 20-mA Setting, 60-µs pulse duration,
        25-Hz sampling rate
  • Clocking using an external or internal clock
  • FIFO with 128-sample depth for ECG and PPG
  • I2C, SPI interfaces: selectable by pin
  • 2.6-mm × 2.1-mm, 0.4-mm Pitch DSBGA package
  • Supplies:
    • Rx: 1.8 V to 1.9 V (LDO bypass),
      2.0 V to 3.6 V (LDO enabled)
    • Tx: 3 V to 5.25 V
    • IO: 1.7 V to Rx_SUP

The AFE4900 device is an analog front-end (AFE) for synchronized electrocardiogram (ECG), photoplethysmogram (PPG) signal acquisition. The device can also be used for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The PPG signal chain supports up to four switching light-emitting diodes (LEDs) and up to three photodiodes (PDs). The LEDs can be switched on using a fully integrated LED driver. The current from the photodiode is converted into voltage by the transimpedance amplifier (TIA) and digitized using an analog-to-digital converter (ADC). The ECG signal chain has an instrumentation amplifier (INA) with a programmable gain that interfaces to the same ADC. A right-leg drive (RLD) amplifier set can be used to the bias for the ECG input pins. AC and dc lead-off detect schemes are supported. The ADC codes from the PPG and ECG phases can be stored in a 128-sample first in, first out (FIFO) block and read out using either an I2C or a serial programming interface (SPI) interface.

The AFE4900 device is an analog front-end (AFE) for synchronized electrocardiogram (ECG), photoplethysmogram (PPG) signal acquisition. The device can also be used for optical bio-sensing applications, such as heart-rate monitoring (HRM) and saturation of peripheral capillary oxygen (SpO2). The PPG signal chain supports up to four switching light-emitting diodes (LEDs) and up to three photodiodes (PDs). The LEDs can be switched on using a fully integrated LED driver. The current from the photodiode is converted into voltage by the transimpedance amplifier (TIA) and digitized using an analog-to-digital converter (ADC). The ECG signal chain has an instrumentation amplifier (INA) with a programmable gain that interfaces to the same ADC. A right-leg drive (RLD) amplifier set can be used to the bias for the ECG input pins. AC and dc lead-off detect schemes are supported. The ADC codes from the PPG and ECG phases can be stored in a 128-sample first in, first out (FIFO) block and read out using either an I2C or a serial programming interface (SPI) interface.

다운로드
추가 정보 요청

전체 데이터시트 및 기타 설계 리소스는 NDA에 따라 제공됩니다. 지금 요청

관심 가지실만한 유사 제품

open-in-new 대안 비교
다른 핀 출력을 지원하지만 비교 대상 장치와 동일한 기능.
AFE4950 활성 웨어러블, 광학 심박수 모니터링 및 SpO2를 위한 초소형 통합 AFE ECG+PPG AFE with integrated high-pass, low-pass filters optimized for dry electrodes.
AFE4960P 활성 일체형 호흡 및 속도 감지 장치가 포함된 2채널 ECG 및 PPG 아날로그 프런트 엔드 ECG+PPG AFE with integrated respiration and pace detection.

기술 자료

검색된 결과가 없습니다. 검색어를 지우고 다시 시도하세요.
1개 모두 보기
상위 문서 유형 직함 형식 옵션 최신 영어 버전 다운로드 날짜
* 데이터 시트 AFE4900 Ultra-low Power, Integrated AFE for Wearable Optical, Electrical Bio-sensing with FIFO datasheet (Rev. B) PDF | HTML 2019. 7. 30

주문 및 품질

지원 및 교육

TI 엔지니어의 기술 지원을 받을 수 있는 TI E2E™ 포럼

콘텐츠는 TI 및 커뮤니티 기고자에 의해 "있는 그대로" 제공되며 TI의 사양으로 간주되지 않습니다. 사용 약관을 참조하십시오.

품질, 패키징, TI에서 주문하는 데 대한 질문이 있다면 TI 지원을 방문하세요. ​​​​​​​​​​​​​​