SNAS758A February 2025 – June 2025 HDC3120
PRODUCTION DATA
The table below illustrates a typical heater activation cycle, assuming a thermal response time (τ99.9%) of 4s, a 1s hold time to allow for moisture evaporation or contaminant removal, and a 30s cool-down period to verify the sensor has fully re-equilibrated with ambient conditions before resuming accurate RH/T measurements.
|
Time (ms) |
Action |
|---|---|
|
0 |
Final ambient RH/T reading completed |
|
+1 |
HEAT_EN = High (1ms signal rise time) |
|
+130 |
Heater current begins |
|
+130 to +5,130 |
Heater hold (4 second warm-up + 1 sec hold) |
|
+5,130.45 |
HEAT_EN = Low (heater turns off in 0.45ms) |
|
+35,130.45 |
RH/T readings resume (after 30 sec cool-down) |