SNAS758A February 2025 – June 2025 HDC3120
PRODUCTION DATA
Refer to the PDF data sheet for device specific package drawings
If the sensor experiences drift due to high humidity and temperature condition or chemical exposure, a targeted recovery process can help:
Bake at elevated temperature and low humidity (100°C at <5% RH) for 5–10 hours, this can help accelerate evaporation and hence enable removal of certain contaminants.
Rehydrate using one of the recommended re-hydration profiles described in Section 8.4, which can help bring the polymer to equilibrium, aiding in the recovery of baseline performance.
Verify the sensor output to confirm restoration of accuracy.
In certain cases, such as those involving corrosive chemicals or physical damage, full recovery is not always possible.