AFE49I30
- Supports signal acquisition of up to 24 phases
- Flexible allocation of 6 LED, 4 PD in each PPG phase
- Synchronized PPG/ECG signal acquisition with each phase configurable as PPG or ECG, and at different data rates
- Synchronized PPG/ECG supports up to 1 kHz data rate
- ECG-only mode supports up to 4 kHz data rate
- PPG signal chain:
- System SNR up to 108-dB at 16 µA PD current
- Low current for continuous operation on a wearable device with a typical value: 15 µA for an LED, 20 µA for the receiver
- PPG Transmitter:
- Supports common anode LED configuration
- 8-Bit programmable LED current with range adjustable from 25 mA to 250 mA
- Mode to fire two LEDs in parallel with independent per-phase current control
- Programmable LED on-time per-phase
- Simultaneous support of 6 LEDs for SpO2, HRM, or multi-wavelength HRM
- PPG Receiver:
- Supports 4 time-multiplexed photodiode inputs
- 2 parallel receivers to sample two PD channels simultaneously
- Individual ambient offset subtraction DAC at each TIA Input with 8-bit per-phase control and range adjustable up to 255-µA
- Individual LED offset subtraction DAC at each TIA input with 5-bit per-phase control and
15.5-µA range - Close to 100 dB ambient rejection up to 10 Hz
- Noise filtering with programmable bandwidth
- Transimpedance Gain: 3.7 kΩ to 1 MΩ
- ECG Signal Chain
- 1-lead ECG with RLD output for bias
- INA gain programmable from 1.6 – 13
- <1 µVrms input noise (in a 1-150 Hz bandwidth) at 4-kHz sample rate
- Integrated 450 Hz Low Pass Filter
- AC, DC lead-off detect with 3.125-100 nA current and programmable interrupt
- Supports external clock and internal oscillator modes
- Option to acquire data synchronized with a system master clock
- Automatic cancellation of DC from ambient, LED
- FIFO with 128-sample depth
- I2C interface
- 2.6-mm × 2.1-mm DSBGA, 0.4-mm pitch
- Supplies: Rx: 1.7 V-1.9 V (LDO bypass); 1.9 V-3.6 V (LDO enabled), Tx: 3 V-5.5 V, IO: 1.7-3.6 V
- Supports operation with 2 supplies - 1.8 V, 5 V
The AFE49I30 is an analog front-end for optical bio-sensing/ PPG and ECG measurement applications. The device supports up to six switching light-emitting diodes (LEDs) and up to four photodiodes. Up to 24 signal phases can be defined and the PPG or ECG signals can be acquired from each phase in a synchronized manner.
索取更多資訊
在 NDA 下提供完整資料表及其他設計資源。立即索取
技術文件
類型 | 標題 | 日期 | ||
---|---|---|---|---|
* | Data sheet | AFE49I30 Ultra-Small, Integrated AFE With FIFO for Multi-Sensor Wearable, Optical Heart-Rate Monitoring and Bio-sensing datasheet | PDF | HTML | 2020年 4月 13日 |
設計與開發
如需其他條款或必要資源,請按一下下方的任何標題以檢視詳細頁面 (如有)。
AFE44S30EVM — AFE4430 整合式 AFE,適用於採用 FIFO 和 SPI 評估模組進行可穿戴式光學生物感測
AFE44S30 評估模組 (EVM) 是評估 AFE44S30 裝置的平台,其為光學生物感測應用領域的類比前端 (AFE),如心率監控 (HRM) 與周邊微血管氧飽和度 (SpO2)。 AFE44S30 支援八個切換發光二極體 (LED) 以及四個光二極體。光二極體電流透過轉阻放大器 (TIA) 轉換為電壓,並使用類比轉數位轉換器 (ADC) 進行數位化。ADC 程式碼可以儲存在具有可編程深度的 128 取樣 FIFO 中。可以對 FIFO 深度進行分區以適應需要儲存的相位。可以使用序列周邊介面 (SPI) 讀取 FIFO。
AFE44S30 也配備具 8 (...)
AFE49I30EVM — 適用於穿戴式光學、電生物感測的 AFE49I30 超低功耗整合式 AFE 評估模組
AFE49I30 評估模組 (EVM) 是一個用於加速和評估與 AFE49I30 產品相關的系統開發活動的平台。AFE49I30 為光學生物感測應用領域的類比前端 (AFE),如心率監控 (HRM) 與周邊微血管含氧飽和度 (SpO2)。
AFE49I30 支援多達六個切換發光二極體 (LED) 和四個光二極體。光二極體電流透過轉阻放大器 (TIA) 轉換為電壓,並使用類比轉數位轉換器 (ADC) 進行數位化。ADC 程式碼可以儲存在具有可編程深度的 128 取樣 FIFO 中。可以對 FIFO 深度進行分區以適應需要儲存的相位。可以使用序列周邊介面 (SPI) 讀取 FIFO。
AFE (...)
BSECUR-3P-ALGORITHMS — B SECUR ECG 演算法和電極設計服務
PSPICE-FOR-TI — PSpice® for TI 設計與模擬工具
封裝 | 針腳 | CAD 符號、佔位空間與 3D 模型 |
---|---|---|
DSBGA (YZ) | 30 | Ultra Librarian |
訂購與品質
- RoHS
- REACH
- 產品標記
- 鉛塗層/球物料
- MSL 等級/回焊峰值
- MTBF/FIT 估算值
- 材料內容
- 認證摘要
- 進行中持續性的可靠性監測
- 晶圓廠位置
- 組裝地點
建議產品可能具有與此 TI 產品相關的參數、評估模組或參考設計。